Back to Blog
How to Calibrate Tablet Friability Testers and Key Parameters
Calibration

How to Calibrate Tablet Friability Testers and Key Parameters

Raise Lab Equipment
December 28, 2024
5 min read

Proper calibration of tablet friability testers ensures accurate assessment of tablet mechanical strength according to pharmacopeial requirements.

Key Calibration Parameters

Rotation Speed

The drum must rotate at a fixed speed of 25 ± 1 rpm:

  • Use a calibrated tachometer to measure speed
  • If speed deviates, adjust motor controller
  • Verify at beginning and during rotation cycle

Fall Height

The tablets must fall from a consistent height:

  • Measured using a scale or gauge
  • Must align with 6.3 ± 0.2 mm standard
  • Check drum interior for wear or damage

Rotation Count

Standard test requires exactly 100 rotations:

  • Verify counter accuracy
  • Test timer correlation (4 minutes at 25 rpm)
  • Check for consistent counting

Drum Condition

Inspect drum for:

  • Cracks or chips
  • Internal surface smoothness
  • Proper mounting and balance

Calibration Procedure

  1. Visual Inspection: Check drum and drive system
  2. Speed Verification: Measure rotation speed with tachometer
  3. Count Verification: Confirm accurate rotation counting
  4. Height Check: Verify fall height specification
  5. Timer Verification: Compare to calibrated stopwatch
  6. Documentation: Record all measurements

Acceptance Criteria

Per USP <1216>:

  • Rotation speed: 25 ± 1 rpm
  • Test duration: 100 rotations
  • Weight loss limit: Typically <1%

Maintenance

  • Keep drum clean and free of tablet residue
  • Check drive belt tension regularly
  • Verify interlock safety systems
  • Clean and inspect regularly