Back to Blog
Dissolution Tester Calibration and Validation Process
Calibration

Dissolution Tester Calibration and Validation Process

Raise Lab Equipment
November 26, 2024
7 min read

Dissolution testers must undergo regular calibration and validation to verify equipment functions correctly and meets regulatory requirements.

Performance Verification Testing (PVT)

PVT is conducted to confirm that the dissolution tester provides consistent and reproducible results using USP Prednisone and Salicylic Acid Tablets as reference standards. Prednisone tablets are commonly used due to their sensitivity to hydrodynamic variations.

USP Methods Covered

Calibration procedures apply to various USP apparatus methods:

  • USP Method 1: Basket apparatus
  • USP Method 2: Paddle apparatus
  • USP Method 5: Paddle over disk
  • USP Method 6: Cylinder apparatus

Key Calibration Parameters

Temperature Control

The dissolution medium must be maintained at 37°C ± 0.5°C, verified using calibrated thermometers or temperature probes.

Rotation Speed

Paddle or basket rotation must be verified at specified speeds (typically 50-100 RPM) using a calibrated tachometer.

Vessel Centering

The distance from the vessel center to the shaft must be within specified tolerances to ensure proper hydrodynamics.

Wobble Testing

Shaft wobble must not exceed specified limits to prevent inconsistent dissolution results.

Regulatory Compliance

Regulatory agencies such as the FDA, USP, and ICH provide guidelines ensuring test results are acceptable for regulatory submissions. Regular calibration is essential for GMP compliance.

Documentation Requirements

All calibration activities must be documented with:

  • Calibration date and next due date
  • Equipment identification
  • Standards used
  • Results and pass/fail criteria
  • Technician signature
R

Raise Lab Equipment

Author and contributor to RaiseLabs blog. Expert in laboratory equipment and quality control standards.